Skip to main navigation Skip to search Skip to main content

Adaptive control of induced hypotension with optimum drug infusion

  • University of Texas at Arlington

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

The authors describe an adaptive control strategy that is suited for automatic control of mean arterial blood pressure (MABP). The performance of the control was evaluated for application of sodium nitroprusside, a vasodilator drug, using a computer simulation with a model of the patient response derived from actual patient data. One feature of the controller is its ability to optimize the quantity of infused drug without generating an offset in the controlled pressure. A comparison with another control strategy which has been used in the clinics is presented. The proposed controller appears to be excellent in regulating MABP.

Original languageEnglish
Title of host publicationBiomedical Engineering Perspectives
Subtitle of host publicationHealth Care Technologies for the 1990's and Beyond
PublisherPubl by IEEE
Pages934-935
Number of pages2
Editionpt 3
ISBN (Print)0879425598
StatePublished - 1990
Externally publishedYes
EventProceedings of the 12th Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Philadelphia, PA, USA
Duration: 1 Nov 19904 Nov 1990

Publication series

NameProceedings of the Annual Conference on Engineering in Medicine and Biology
Numberpt 3
ISSN (Print)0589-1019

Conference

ConferenceProceedings of the 12th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
CityPhiladelphia, PA, USA
Period1/11/904/11/90