Skip to main navigation Skip to search Skip to main content

Biomechanical investigation of two long bone growth modulation techniques by finite element simulations

  • Manuel Schneider
  • , Jan Buschbaum
  • , Alexander Joeris
  • , Oliver Röhrle
  • , Jonathan Dwyer
  • , James B. Hunter
  • , Richard A.K. Reynolds
  • , Theodor F. Slongo
  • , Boyko Gueorguiev
  • , Peter Varga
  • AO Research Institute
  • AO Clinical Investigation and Documentation
  • University of Stuttgart
  • University Hospital of North Staffordshire
  • Nottingham University Hospitals NHS Trust
  • Alfred I. duPont Hospital for Children
  • University of Bern

Research output: Contribution to journal › Article › peer-review

7 Scopus citations

Abstract

Implants used to correct pathological varus–valgus deformities (VVD) and leg length discrepancies (LLD) may not be optimized for the specific treatment, as suggested by their off-label use. Detailed analysis of this issue has been limited by the poorly understood mechanical behavior of the growing physis and ignorance of the loads acting on the implants. The aim of this study was to predict and compare the loading conditions of a growth modulation implant in VVD and LLD treatments. Idealized finite element (FE) models of the juvenile distal femur treated with the Eight-Plate implant were developed for VVD and LLD. Bone growth was simulated using thermal strains. The axial force in the plate was compared between the two treatments. Case-specific plate forces were predicted by virtually reproducing the screw deformation visible on radiographs of LLD (N = 4) and VVD (N = 4) clinical cases. The simple FE models reproduced the clinical implant deformations well. The resulting forces ranged from 129 to 580 N for the VVD patients. For LLD, this range was from 295 to 1002 N per plate, that is, 590–2004 N for the entire physis. The higher forces in LLD could be explained by restricted screw divergence in the double-sided implant application. For the first time, the loading conditions of a growth modulation implant were investigated and compared between two treatments by FE analyses, and the range of case-specific loads was predicted. These simulation tools may be utilized for guiding appropriate usage and for efficient development of implants.

Original languageEnglish
Pages (from-to)1398-1405
Number of pages8
JournalJournal of Orthopaedic Research
Volume36
Issue number5
DOIs
StatePublished - May 2018
Externally publishedYes

Keywords

  • Eight-Plate
  • bone growth modulation
  • finite element analysis
  • leg length discrepancy
  • varus–valgus deformity