Abstract
We report the successful closure of an extracardiac Fontan fenestration with a bio-absorbable device, which may be refenestrated by a transcatheter route when needed, in a 10-year-old boy. The patient presented with cyanosis two years after an extracardiac Fontan operation. Echocardiography revealed a moderate shunt from the Fontan circulation into the systemic circulation with a mean pressure gradient of 3-4 mmHg. Treadmill testing revealed a significant decrease in oxygen saturation (down to the low 50's from a baseline level of 80-85%). Cardiac catheterization revealed normal pressure in the Fontan circuit. A temporary balloon occlusion test showed that the defect was suitable for permanent occlusion. The fenestration was then occluded by a bio-absorbable Biostar (NMT medical, Boston, USA) atrial septal occluder device. The oxygen saturation on room air increased up to 95% after closure.
| Translated title of the contribution | Successful transcatheter closure of a Fontan fenestration with a bioabsorbable Biostar occluder |
|---|---|
| Original language | Turkish |
| Pages (from-to) | 623-627 |
| Number of pages | 5 |
| Journal | Turk Kardiyoloji Dernegi Arsivi |
| Volume | 40 |
| Issue number | 7 |
| DOIs | |
| State | Published - 2012 |
| Externally published | Yes |
Keywords
- Child
- Cyanosis
- Fontan procedure/adverse effects
- Heart catheterization/ instrumentation
- Heart defects, congenital
Fingerprint
Dive into the research topics of 'Successful transcatheter closure of a Fontan fenestration with a bioabsorbable Biostar occluder'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver