Rebound tonometry by iCare 200 (IC200): comparison with Tono-Pen in the operating room and Goldmann applanation in the clinic

  • Allison C. Umfress
  • , Tanya S. Glaser
  • , Pimpiroon Ploysangam
  • , Sharon F. Freedman
  • Duke University

Research output: Contribution to journal › Article › peer-review

10 Scopus citations

Abstract

Background: Tonometry guides decision making in children with known or suspected glaucoma. The iCare IC200 is a recently FDA-approved rebound tonometer that can measure intraocular pressure (IOP) in both supine and upright patients. The purpose of this study was to evaluate this new tonometer against more commonly used standard instruments: Tono-Pen in the operating room (OR arm) and Goldmann applanation in the clinic (clinic arm). Methods: The OR arm included sequential children undergoing general anesthesia. The clinic arm included cooperative children seen in clinic. IOP was measured twice (right eye then left eye, repeat) with IC200 and either Tono-Pen (OR arm) or Goldmann (clinic arm). Results: A total of 99 children were included: 75 for the OR arm and 24 for the clinic arm. There was a strong correlation between the IOP measured by IC200 and each device (R2 = 0.81 for Tono-Pen; R2 = 0.78 for Goldmann). In the OR arm, the overall mean IOP difference (ΔIOP [Tono-Pen – IC200]) was −1.6 mm Hg. For eyes with corneal edema (n = 12), ΔIOP ranged from −8 to 15 mm Hg. At IOP >20 mm Hg, mean ΔIOP was −3.7 mm Hg. In the clinic arm, the mean ΔIOP (ΔIOP [Goldmann – IC200]) was −1.2 mm Hg. Conclusions: In eyes of children under general anesthesia with IOP <20 mm Hg without corneal edema, IC200 and Tono-Pen correlated well. In cooperative clinic patients, Goldmann and IC200 correlated well over the range of IOP studied.[Formula

Original languageEnglish
Pages (from-to)329.e1-329.e6
JournalJournal of AAPOS
Volume25
Issue number6
DOIs
StatePublished - Dec 2021
Externally publishedYes

Fingerprint

Dive into the research topics of 'Rebound tonometry by iCare 200 (IC200): comparison with Tono-Pen in the operating room and Goldmann applanation in the clinic'. Together they form a unique fingerprint.

Cite this