Abstract
Objective: To compare the work of breathing (WOB) in premature neonates supported with high-flow nasal cannula (HFNC) and nasal continuous positive airway pressure (NCPAP). Study design: Eighteen preterm neonates <2.0 kg on HFNC or NCPAP support were studied in a random order. A ventilator was used to deliver 6 cm H2O of NCPAP with nasal prongs. High-flow nasal cannula delivered with Vapotherm (VAPO) at 3, 4 and 5l/min was used. Tidal ventilation was obtained using respiratory inductance plethysmography calibrated with face-mask pneumotachography. An esophageal balloon estimated pleural pressure from which changes in end distending pressure were calculated. Inspiratory, elastic and resistive WOB and respiratory parameters were calculated. Results: No differences were found in the WOB for all settings. Changes in end distending pressure did not vary significantly over all device settings except VAPO at 5l/min. Conclusion: In these preterm infants with mild respiratory illness, HFNC provided support comparable to NCPAP.
| Original language | English |
|---|---|
| Pages (from-to) | 476-480 |
| Number of pages | 5 |
| Journal | Journal of Perinatology |
| Volume | 26 |
| Issue number | 8 |
| DOIs | |
| State | Published - Aug 2006 |
| Externally published | Yes |
Keywords
- High-flow nasal cannula
- Nasal continuous positive airway pressure
- Vapotherm
- Work of breathing