Resumen
Stent angioplasty of patent ductus arteriosus has been shown to be a viable alternative to operative shunt placement in cyanotic neonates. With broader implementation of this strategy, novel complications are bound to arise. We present a series of cases evaluated for ductal stent angioplasty in which a dilated and torturous ductus arteriosus compressed the left mainstem bronchus. After reviewing our recent experience with ductal stenting and isolated Blalock-Taussig shunts, our best estimate of the incidence of bronchial compression by the dilated ductus is 4.6% (3/64, 95% confidence interval 1.0–12.9%). Awareness of the airway and other nonvascular contents of the thorax is an important consideration prior to ductal stenting.
| Idioma original | English |
|---|---|
| Páginas (desde-hasta) | 1158-1162 |
| Número de páginas | 5 |
| Publicación | Catheterization and Cardiovascular Interventions |
| Volumen | 95 |
| N.º | 6 |
| DOI | |
| Estado | Published - 1 may 2020 |