Cross-vendor validation of liver magnetic resonance elastography

  • Cincinnati Children's Hospital Medical Center
  • Mayo Clinic Rochester, MN
  • Koninklijke Philips N.V.

Producción científica › revisión exhaustiva

62 Citas (Scopus)

Resumen

Purpose: To evaluate and validate the reproducibility of MR Elastography (MRE)-derived liver stiffness values on two different MR vendor platforms performed on the same subject on the same day. Methods: This investigation was approved by the hospital IRB. MRE exams were performed twice in identical fashion in eight volunteers and in five clinical patients on two different 1.5 T MR scanners—once on a Philips MR scanner and immediately afterward in back-to-back fashion on a General Electric MR scanner, or vice versa. All scan parameters were kept identical on the two platforms to the best extent possible. After the MRE magnitude and phase images were obtained, the data were converted into quantitative images displaying the stiffness of the liver parenchyma. Mean liver stiffness values between the two platforms were compared using interclass correlation with a p value <0.05 considered statistically significant. Results: Interclass correlation coefficient (ICC) value of 0.994 was obtained for 13 subjects with p value <0.001 indicating a significantly positive correlation. Conclusion: As MRE gains in acceptance and as its availability becomes more widespread, it is important to ascertain and confirm that liver stiffness values obtained on different MRE vendor platforms are consistent and reproducible. In this small pilot investigation, we demonstrate that liver stiffness measurement with MRE is reproducible and has very good consistency across two vendor platforms.

Idioma originalEnglish
Páginas (desde-hasta)789-794
Número de páginas6
PublicaciónAbdominal Imaging
Volumen40
N.º4
DOI
EstadoPublished - 5 dic 2014

Huella

Profundice en los temas de investigación de 'Cross-vendor validation of liver magnetic resonance elastography'. En conjunto forman una huella única.

Citar esto