Ir directamente a la navegación principal Ir directamente a la búsqueda Ir directamente al contenido principal

Design and testing of a novel slot retractor for neonatal frenotomy

  • Tariq Rahman
  • , Nghia N. Tran
  • , Gita Jani
  • , Chinedum N. Udenze-Utah
  • , Shehla S. Siddiqui
  • , Leonard M. Golub
  • , Adriana Restrepo-Ormsby
  • , Thomas H. Shaffer
  • Alfred I. duPont Hospital for Children
  • Drexel University
  • Doylestown Hospital

Producción científica › revisión exhaustiva

1 Cita (Scopus)

Resumen

A new slot retractor for frenotomy has been developed that offers a unique design that will lead to fewer complications and allow primary care physicians (neonatologists, pediatricians, family physicians) to perform frenotomy in the nursery or office. The novel 3-dimensional printed retractor has side walls for better tissue protection, a built-in slot, and an ergonomic handle. An efficacy and safety study was conducted with 43 infants with the newly designed slot retractor. Results showed that there were no partial divisions at discharge; all babies were able to breastfeed almost immediately after frenotomy and continued to breastfeed until hospital discharge, and there were no adverse events.

Idioma originalEnglish
Páginas (desde-hasta)186-190
Número de páginas5
PublicaciónJournal of Clinical Engineering
Volumen41
N.º4
DOI
EstadoPublished - 1 oct 2016

Huella

Profundice en los temas de investigación de 'Design and testing of a novel slot retractor for neonatal frenotomy'. En conjunto forman una huella única.

Citar esto