Hemoglobin A1c and mean glucose in patients with type 1 diabetes: Analysis of data from the juvenile diabetes research foundation continuous glucose monitoring randomized trial

  • Darrell M. Wilson
  • , Dongyuan Xing
  • , Roy W. Beck
  • , Jennifer Block
  • , Bruce Bode
  • , Larry A. Fox
  • , Irl Hirsch
  • , Craig Kollman
  • , Lori Laffel
  • , Katrina J. Ruedy
  • , Michael Steffes
  • , William V. Tamborlane
  • Stanford University
  • Jaeb Center for Health Research
  • Atlanta Diabetes Associates
  • Alfred I. duPont Hospital for Children
  • University of Washington
  • Joslin Diabetes Center
  • University of Minnesota Twin Cities
  • Yale University

Producción científica › revisión exhaustiva

61 Citas (Scopus)

Resumen

OBJECTIVE - To determine the relationship between mean sensor glucose concentrations and hemoglobin A1c (HbA1c) values measured in the Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications laboratory at the University of Minnesota in a cohort of subjects with type 1 diabetes from the Juvenile Diabetes Research Foundation continuous glucose monitoring randomized trial. RESEARCH DESIGN AND METHODS - Near-continuous glucose sensor data (≥4 days/week) were collected for 3 months before a central laboratory-measured HbA1c was performed for 252 subjects aged 8-74 years, the majority of whom had stable HbA 1c values (77% within 60.4% of the patient mean). RESULTS - The slope (95% CI) for mean sensor glucose concentration (area under the curve) versus a centrally measured HbA1c was 24.4 mg/dL (22.0-26.7) for each 1% change in HbA1c, with an intercept of -16.2 mg/dL (-32.9 to 0.6). Although the slope did not vary with age or sex, there was substantial individual variability, with mean sensor glucose concentrations ranging from 128 to 187 mg/dL for an HbA1c of 6.9-7.1%. The root mean square of the errors between the actual mean sensor glucose concentration versus the value calculated using the regression equation was 14.3 mg/dL, whereas the median absolute difference was 10.1 mg/dL. CONCLUSIONS - There is substantial individual variability between the measured versus calculated mean glucose concentrations. Consequently, estimated average glucose concentrations calculated from measured HbA1c values should be used with caution.