The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes

  • Darrell M. Wilson
  • , Roy W. Beck
  • , William V. Tamborlane
  • , Mariya J. Dontchev
  • , Craig Kollman
  • , Peter Chase
  • , Larry A. Fox
  • , Katrina J. Ruedy
  • , Eva Tsalikian
  • , Stuart A. Weinzimer
  • , Rosanna Fiallo-Scharer
  • , Laurel Messer
  • , Barbara Tallant
  • , Michael J. Tansey
  • , Linda F. Larson
  • , Julie Coffey
  • , Joanne Cabbage
  • , Tim Wysocki
  • , Nelly Mauras
  • , Keisha Bird
  • Kim Englert, Bruce A. Buckingham, Jennifer M. Block, Paula Clinton, Kimberly Caswell, Elizabeth A. Doyle, Heather Mokotoff, Amy Steffen, Dongyuan Xing, Cynthia R. Stockdale, Judy Jackson, Michael W. Steffes, Jean M. Bucksa, Maren L. Nowicki, Carol A. Van Hale, Vicky Makky, Gilman D. Grave, Mary Horlick, Karen Teff, Karen K. Winer, Dorothy M. Becker, Patricia Cleary, Christopher M. Ryan, Neil H. White, Perrin C. White
  • Jaeb Center for Health Research
  • Stanford University
  • Yale University
  • Alfred I. duPont Hospital for Children
  • University of Colorado Denver
  • University of Iowa
  • University of Minnesota Central Laboratory
  • National Institutes of Health
  • Data and Safety Monitoring Board

Producción científicarevisión exhaustiva

116 Citas (Scopus)

Resumen

OBJECTIVE - To evaluate the accuracy and precision of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. RESEARCH DESIGN AND METHODS - In 30 children with type 1 diabetes (mean age 11.2 ± 4.1 years), the Navigator glucose values were compared with reference serum glucose values of blood samples obtained in an inpatient clinical research center and measured in a central laboratory using a hexokinase enzymatic method and in an outpatient setting with a FreeStyle meter. Median absolute difference (AD) and median relative absolute difference (RAD) were computed for sensor-reference and sensor-sensor pairs. RESULTS - The median AD and RAD were 17 mg/dl and 12%, respectively, for 1,811 inpatient sensor-reference pairs and 20 mg/dl and 14%, respectively, for 8,639 outpatient pairs. The median RAD between two simultaneous Navigator measurements (n = 1,971) was 13%. Ninety-one percent of sensors in the inpatient setting and 81% of sensors in the outpatient setting had a median RAD ≤20%. CONCLUSIONS - The Navigator's accuracy does not yet approach the accuracy of current-generation home glucose meters, but it is sufficient to believe that the device has the potential to be an important adjunct to treatment of youth with type 1 diabetes.

Idioma originalEnglish
Páginas (desde-hasta)59-64
Número de páginas6
PublicaciónDiabetes Care
Volumen30
N.º1
DOI
EstadoPublished - ene 2007