Resumen
Study Design.: Retrospective review. Objective.: To report on the safety of repetitive transcranial electric stimulation (RTES) for eliciting motor-evoked potentials during spine surgery. Summary of Background Data.: Theoretical concerns over the safety of RTES have hindered broader acceptance of transcranial electric motor-evoked potentials (tceMEP), despite successful implementation of spinal cord monitoring with tceMEPs in many large spine centers, as well as their apparent superiority over mixed-nerve somatosensory-evoked potentials (SSEP) for detection of spinal cord injury. Methods.: The records of 18,862 consecutive patients who met inclusion criteria and underwent spine surgery with tceMEP monitoring were reviewed for RTES-related complications. Results.: This large retrospective review identified only 26 (0.14%) cases with RTES-related complications; all but one of these were tongue lacerations, most of which were self-limiting. Conclusions.: The results demonstrate that RTES is a highly safe modality for monitoring spinal cord motor tract function intraoperatively.
| Idioma original | English |
|---|---|
| Páginas (desde-hasta) | 1046-1049 |
| Número de páginas | 4 |
| Publicación | Spine |
| Volumen | 36 |
| N.º | 13 |
| DOI | |
| Estado | Published - 1 jun 2011 |
Huella
Profundice en los temas de investigación de 'Transcranial electric motor evoked potential monitoring during spine surgery: Is it safe?'. En conjunto forman una huella única.Citar esto
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver